I-E67-TB82FF-2A February 13, 2002
31
CONDUCTIVITY/CONCENTRATION TRANSDUCER BLOCK (cont’d)
IDX Name Description Bytes ST AR Default Valid Range Note
49 TEMP_ CAL_UNIT
Temperature
Calibration Value
Units
2 S R/W 1001 (°C) Table 4
50 PV_CAL_SLOPE PV Calibration Slope 4 S R/W 100% 20 to 500 %
51
PV_CAL_SLOPE_
UNIT
PV Calibration Slope
Unit
2 N R 1342 (%) 1342 (%)
52 PV_CAL_OFFSET PV Calibration Offset 4 S R/W
0 µS/cm
-20 to +20 µS/cm
(Sensor Group A)
-4 to +4 µS/cm
(Sensor Group B)
-0.8 to +0.8 µS/cm
(Sensor Group C)
53
PV_CAL_
OFFSET_UNIT
PV Calibration Offset
Unit
2NR
1586 (µS/cm)
Table 2
54 PV_CAL_RESET Reset PV Calibration 1 S R/W 0 (OFF)
0 – OFF
1 – ON
55
TEMP_CAL_
SLOPE
Temperature
Calibration Slope
4 S R/W 100% 20 to 150 %
56
TEMP_CAL_
SLOPE_UNIT
Temperature
Calibration Slope
Unit
2 N R 1342 (%)
57
TEMP_CAL_
OFFSET
Temperature
Calibration Offset
4SR/W 0 °C
-40 to +40 °C
(3K Balco, PT100)
-100 to +100 °C
(4.75K RTD)
58
TEMP_CAL_
OFFSET_UNIT
Temperature
Calibration Offset
Unit
2 S R/W 1001 (°C) Table 4
59
TEMP_CAL_
RESET
Reset Temperature
Calibration
1 S R/W 0 (OFF)
0 – OFF
1 – ON
60
LCD_
SECONDARY
LCD Secondary
Display Option
1SR/W
0 (Temperature in
°C)
Table 5
61 SECURITY_LOCK
Security Menu
Password-enable for
Keypad Operation
1 S R/W 0 (DISABLE)
0 – Disable
1- Enable
62
CONFIGURATION
_LOCK
Configuration Menu
Password-enable for
Keypad Operation
1 S R/W 0 (DISABLE)
0 – Disable
1 - Enable
63
CALIBRATION_
LOCK
Calibration Menu
Password-enable for
Keypad Operation
1 S R/W 0 (DISABLE)
0 – Disable
1- Enable
64
SECURITY_
PASSWORD
3-digit Password for
Password-enabled
Keypad Operations
3 S R/W 000 000 to 999
65
EXTENDED_
STATUS
Extended Status 9 D R